Changes in the Glow Curve of Thermoluminescent Dosimeter According to Magnetic Field Strength

Journal of Magnetics, Volume 27, Number 4, 31 Dec 2022, Pages 468-473
Jeong-Ho Kim (Sunlin University), Man-Seok Han * (Kangwon National University), Min Cheol Jeon * (Daejeon Health Institute of Technology)
Abstract
The thermoluminescent dosimeter is the most commonly used instrument for measuring radiation dose. Based
on the previous study that the glow curve changes in the magnetic field environment of a thermal luminescence
dose device, the change of the glow curve according to the change of the magnetic field strength is evaluated.
Accordingly, the change in the glow curve was analyzed by irradiating 0.5-2 cGy to the thermoluminescent
chips with a radiation generator while changing the magnetic field strength using a permanent magnet. As a
result of the evaluation, an average increase of 0.3% was observed at 90~120 ℃, an average decrease of -10.6%
at 120-160 ℃, and an average increase of 3.2% at 160-260 ℃. Therefore, when applying the thermoluminescent
dosimeter, it is necessary to evaluate the dose considering the error rate according to the magnetic field
strength.
Keywords: thermal luminescence dosimeter; glow curve; magnetic field strength; exothermic temperature; dose evaluation