Analysis of Magnetic Resonance Signals from Diffusion Weighted Imaging using Compressed Sensitivity Encoding Technique

Journal of Magnetics, Volume 24, Number 2, 30 Jun 2019, Pages 217-222
Ji-Sung Jang (Department of Radiology, Asan Medical Center), Yong-Soo Han (Department of Radiological Science, Hallym Polytechnic University), Mi-Ae Jeong * (Department of Dental Hygiene, Kangwon National University)
Abstract
The purpose of this study was to compare image quality between compressed sensitivity encoding (CS) and sensitivity encoding (SENSE) in single-shot turbo spin echo diffusion-weighted imaging (TSE-DWI). Signal-tonoise ratios (SNR), apparent diffusion coefficient (ADC) values, and geometric accuracies were measured with an American College of Radiology head phantom. CS showed a high mean SNR, less variation in ADC values, and reduced imaging acquisition time (21.7 %-39.3 %) compared with SENSE. In addition, there was no significant difference in geometric accuracy between SENSE and CS. In conclusion, in comparison with SENSE TSE-DWI, CS TSE-DWI resulted in a reduced imaging acquisition time and improved SNR, while maintaining image quality.
 
Keywords: magnetic resonance imaging; magnetic resonance signal to noise ratio; sensitivity encoding; geometric accuracy
DOI: https://doi.org/10.4283/JMAG.2019.24.2.217