Home 
Retrieve | Search
 
Issues
No.3(pp.67-109)
Home > Issues > Volume 3 (1998) > No.3(pp.67-109)
 
Characteristics of HTS SQUID-based Susceptometer
Journal of Magnetics, Volume 3, Number 3, 30 Sep 1998, Pages 82-85
V. P. Timofeev(Department of Physics, Sun Moon University, Inst. for Low Temp. Phys. & Eng., Nat'l Academy of Sciences of Ukraine), C.G. Kim(Department of Physics, Sun Moon University), V. I. Shnyrkov(Inst. for Low Temp. Phys. & Eng., Nat'l Academy of Sciences of Ukraine)
Abstract
A portable HTS RF SQUID-based system, weighing less than 20 kg has been built for susceptometry applications in weak magnetic fields, It includes a YBCO sensor for measuring the axial magnetic field component with a resolution of about 7×10-13 T/Hz1/2. This is determined by the intrinsic magnetic noise in the quasi-white noise region. There is a relaxation for a sudden increase in field due to magnetic flux creep in HTS. In this instance the time did not exceed 3~5 minutes.
Keywords:
DOI:
Full Text: PDF
 
Copyright(c) 2009 The Korean Magnetics Society All rights reserved. E-mail : komag@unitel.co.kr