Peripheral Blood Flow Velocity and Peripheral Pulse Wave Velocity Measured Using a Clip-type Pulsimeter Equipped with a Permanent Magnet and a Hall Device
Journal of Magnetics, Volume 20, Number 1, 31 Mar 2015, Pages 47-51
Keun-Ho Kim (Department of Oriental Biomedical Engineering, Sangji University), Sang-Suk Lee * (Department of Oriental Biomedical Engineering, Sangji University)
Abstract

We measured radial arterial pulse signals using a prototype of a clip-type pulsimeter equipped with a
permanent magnet and a Hall device, which produced signals through a voltage-detecting circuit. The systolic peak time and the reflective peak time for a temporally pulsed signal were analyzed for an arbitrary pulse wave at one position of a small permanent magnet. The measured value of the peripheral pulse wave velocity was about 1.25-1.52 m/s, demonstrating the accuracy of this new method. To measure the peripheral blood flow velocity, we simultaneously connected the radial artery pulsimeter to a photoplethysmography meter. The average value of the peripheral blood flow velocity was about 0.27-0.50 m/s.


 

Keywords: radial artery pulse wave; clip-type pulsimeter; photoplethysmography; systolic peak time; reflective peak time; peripheral blood flow velocity; peripheral pulse wave velocity
DOI: http://dx.doi.org/10.4283/JMAG.2015.20.1.047