Home 
Retrieve | Search
 
Issues
Retrieve
Home > Issues > Retrieve
 
Optimization of Operation Frequency of Orthogonal Fluxgate Sensor Fabricated with Co Based Amorphous Wire
Journal of Magnetics, Volume 18, Number 2, 30 Jun 2013, Pages 159-162
Young-Hak Kim (Department of Electrical Engineering, Pukyong National University), Yongmin Kim (Department of Information & Communication Engineering, Kyungsung University), Chang-Seob Yang (6th R&D Institute, Agency for Defense Development), Kwang-Ho Shin * (Department of Information & Communication Engineering, Kyungsung University)
Abstract

We present how to optimize the operation condition including frequency of the orthogonal fluxgate sensor in this paper. The orthogonal fluxgate sensor was fabricated with a Co-based amorphous wire with 10 mm long and 100 μm in the diameter and a 270-turn pickup coil wound on the amorphous wire. In order to investigate the frequency dependence of the sensitivity, output spectra of the sensor which was connected by using a coaxial cable with various lengths of 0.5-5 m were measured with a RF lock-in amplifier. The maximum sensitivities were obtained at different frequencies according to coaxial cable lengths. It was found that the optimal operation frequencies, at which maximum sensitivities were appeared, were almost identical to the frequencies of impedance resonance. The maximum sensitivity and optimal operation frequency were 1.1 V/Oe (≈ 11000 V/T) and 1.25 MHz respectively.


 

Keywords: orthogonal fluxgate; CoFeSiB amorphous wire; MHz operation; LC resonance; sensitivity
DOI: http://dx.doi.org/10.4283/JMAG.2013.18.2.159
Full Text: PDF
 
Copyright(c) 2009 The Korean Magnetics Society All rights reserved. E-mail : komag@unitel.co.kr