Analysis of Images According to the Fluid Velocity in Time-of-Flight Magnetic Resonance Angiography, and Contrast Enhancement Angiography
Journal of Magnetics, Volume 19, Number 2, 30 Jun 2014, Pages 185-191
Eng-Chan Kim (Department of Physics, Yeungnam University), Yeong-Cheol Heo (Department of Radiology, Kyung Hee University Hospital at Gang-dong), Jae-Hwan Cho (Department of International Radiological Science, Hallym University of Graduate Studies), Hyun-Jeong Lee (Department of Physics, Yeungnam University), Hae-Kag Lee * (Department of Computer Science and Engineering, Soonchunhyang University)
Abstract

In this study we evaluated that flow rate changes affect the (time of flight) TOF image and contrast-enhanced (CE) in a three-dimensional TOF angiography. We used a 3.0T MR System, a nonpulsatile flow rate model. Saline was used as a fluid injected at a flow rate of 11.4 cm/sec by auto injector. The fluid signal strength, phantom body signal strength and background signal strength were measured at 1, 5, 10, 15, 20 and 25-th cross-section in the experienced images and then they were used to determine signal-to-noise ratio and contrastto- noise ratio. The inlet, middle and outlet length were measured using coronal images obtained through the maximum intensity projection method. As a result, the length of inner cavity was 2.66 mm with no difference among the inlet, middle and outlet length. We also could know that the magnification rate is 49-55.6% in inlet part, 49-59% in middle part and 49-59% in outlet part, and so the image is generally larger than in the actual measurement. Signal-to-noise ratio and contrast-to-noise ratio were negatively correlated with the fluid velocity and so we could see that signal-to-noise ratio and contrast-to-noise ratio are reduced by faster fluid velocity. Signal-to-noise ratio was 42.2-52.5 in 5-25th section and contrast-to-noise ratio was from 34.0-46.1 also not different, but there was a difference in the 1st section. The smallest 3D TOF MRA measure was 2.51 ± 0.12 mm with a flow velocity of 40 cm/s. Consequently, 3D TOF MRA tests show that the faster fluid velocity decreases the signal-to-noise ratio and contrast-to-noise ratio, and basically it can be determined that 3D TOF MRA and 3D CE MRA are displayed larger than in the actual measurement.


 

Keywords: TOF angiography; fluid velocity; length of inner cavity; signal-to-noise ratio; contrast-to-noise ratio
DOI: http://dx.doi.org/10.4283/JMAG.2014.19.2.185
Full Text: PDF
 
Copyright(c) 2009 The Korean Magnetics Society All rights reserved. E-mail : komag@unitel.co.kr